GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to recover.
Source facts
Action
On site visits were performed from 3/6/08 - 12/19/08 to update software on all 5 units.
Initiated
2008-03-06
Root cause
Software design
Terminated
2014-09-04
Distribution
Worldwide Distribution -- USA including states of MD, TX, and VA and country of China.
Recall number
Z-1093-2009
Recall status
Terminated
CDRH record ID
78879
FDA classified
2009-03-12
Recall event number
51159