Open original FDA record ↗Recall record
- FDA identifier
- Z-1093-2017
- Company in FDA record
- Acumedia Manufacturers, Inc.
- FDA record creation date
- Jan 25, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
light precipitate may form in prepared plates of Acumedia Tryptic Soy Agar (7100) Lot# 108166. While there is no change in growth performance of this medium, the formation of particulate matter may be misinterpreted as growth in poured plates thereby leading to inaccurate counts.
Source facts
- Action
- Neogen sent a Recall Notification dated March 8, 2016, to all affected customers on April 22, 2016. via e-mail. The recall notification included a description of the reason for the recall, affected product, and offer for replacement. If any Tryptic Soy Agar from this lot remains in your inventory, and you would like to replace this lot, please contact the customer service team at Neogen Corporation by email at foodsafety@neogen.com or by telephone at (800)-234-5333. We would like to assure you that we are committed to providing products that meet our strict quality standards for your testing needs. Neogen will investigate this isolated incident to determine the root cause and implement measures to ensure that this issue does not happen in the future.
- Initiated
- 2016-04-22
- Root cause
- Component design/selection
- Terminated
- 2018-09-19
- Distribution
- Worldwide Distribution - US Distribution to the states of : AR, CA, CT, FL, GA, ID, IL, IN, MD, ME, MI, NC, ND, NY, OH, OR, TX, TN, UT, VA, WI. and to the countries of : CANADA, UK, ISRAEL, MEXICO, AUSTRALIA, COLOMBIA, TAIWAN, BRAZIL, CHILE and INDONESIA.
- Recall number
- Z-1093-2017
- Recall status
- Terminated
- CDRH record ID
- 152054
- FDA record created
- 2017-01-25
- Recall event number
- 76053