Component of in vitro diagnostic device was incorrectly installed causing incorrect reporting of patient results.
Source facts
Action
The recalling firm notified consignees by letter on 7/21/2003 and advised of the potential for incorrect optical bundle installation. The firm provided an optical test procedure with materials to conduct an on site performance test. Consignees were instructed not to report patient results until optical performance has been verified. Firm will replace or repair failing instruments.
Initiated
2003-07-21
Root cause
Other
Terminated
2003-11-19
Distribution
Product was distributed to domestic and military end users nationwide and international distributors.
Recall number
Z-1094-03
Recall status
Terminated
CDRH record ID
28655
FDA classified
2003-08-20
Recall event number
26884