Open original FDA record ↗Recall record
- FDA identifier
- Z-1094-2013
- Company in FDA record
- Trophy Sas
- FDA classification date
- May 2, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.
Source facts
- Action
- Trophy, A Carestream Dental Company, - A Subsidiary of Carestream Health Inc. The Center for Devices (CDRH) understands that the corrective action plan (CAP) involving 167 units will consist of implementing rework procedure TL008, which describes the procedure for a bending the metallic strip the female connector to insure strong connection with the male counterpart. The CAP appears to adequately address the problem and is hereby approved For further questions please call (585) 781-1997.
- Initiated
- 2012-09-18
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-05-02
- Distribution
- USA Nationwide Distribution
- Recall number
- Z-1094-2013
- Recall status
- Terminated
- CDRH record ID
- 117229
- FDA classified
- 2013-05-02
- Recall event number
- 64589