Class II

Reason stated in the source

Firm discovered a firmware memory leak with a supplier-manufactured component.

Source facts

Action
The firm, Hillrom, issued an "Urgent Medical Device Correction Notice", dated 04/13/2022, to impacted consignees via email. The consignee is instructed to identify if they have the impacted product component, Part P2519NNC3A2 and also distribute the notification to personnel as well ensure that the users monitor for system alerts. Notify Hillrom Technical support if there is a loss of connectivity to the nurse call server. Note, error indications will display on multiple user interfaces. Central nurses' stations will state the room(s) is down (disconnected). In patient room(s), graphical audio stations will display clear message stating Station down. Contact Nurse directly . In addition, standard audio stations will have an LED Error state indicator. Complete and return the response form to william.knouse@hillrom.com. Hillrom is testing a firmware update for the Power over Ethernet (PoE) switch and anticipates availability in April 2022. Upon receipt of the response form, Hillrom will contact you to schedule an update to your system once firmware is available. If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support at +1 919 854 3361 or william.knouse@hillrom.com
Initiated
2022-04-13
Root cause
Software design
Terminated
2023-12-20
Distribution
US Distribution to states of: NC and VA.
Recall number
Z-1094-2022
Recall status
Terminated
CDRH record ID
192992
FDA record created
2022-05-13
Recall event number
89961

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗