Z-1094-2026: Plum Duo Infusion System, List Number 40002-04-01

Recall record Z-1094-2026 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1094-2026
Company in FDA record
ICU Medical, Inc.
FDA classification date
Jan 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION dated 12/15/25 was sent to customers. Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately informed of this notice. If the devices are used at another location, please ensure this communication is delivered to that location. 2. If the primary speaker fails, obtain another pump to continue the infusion. Remove the affected pump from clinical use until the speaker can be replaced. 3. Complete and return the attached Customer Response Form to icumedmarketactionsteam@icumed.com within 10 days of receipt to acknowledge your understanding of this notification. 4. If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to icumedmarketactionsteam@icumed.com. Follow-up Actions by ICU Medical: ICU Medical will address the issue described in this letter through a software update. This update will ensure the backup buzzer will always generate audio for alarms, malfunctions and alerts. ICU Medical will contact you to schedule the software correction for this issue. Device Correction Inquiries https://icumed.custhelp.com/app/market-action For any questions regarding this action
Initiated
2025-12-15
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1094-2026
Recall status
Open, Classified
CDRH record ID
217676
FDA classified
2026-01-15
Recall event number
98129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗