Skip to content

Public FDA record

Z-1095-2012: Inspira AIR Balloon Dilation System, size 18 x 40; Manufactured by Acclarent, Inc. 1525-G O'Brien Drive,

Recall record Z-1095-2012 for Acclarent, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1095-2012
Company in FDA record
Acclarent, Inc.
FDA classification date
Mar 1, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Acclarent received reports of difficulty deflating the balloon during the procedure, which could potentially result in airway obstruction.

Source facts

Action
Acclarent sent an "Urgent Voluntary Product Recall" letter dated January 31, 2012 by US mail, return receipt requested to all affected customers. Additionally, customers known to have scheduled procedures using the affected product were notified by phone to advise them of the recall. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory immediately, discontinue use and return any remaining in stock per the instructions provided. The letter provides additional information and photographs illustrating how to identify the recalled product. Customers were advised to complete the enclosed Business Reply Card regardless of whether they have or do not have product subject to this recall in their possession and mail to Stericycle. Customers were instructed to distribute this information to all staff within their department who uses the affected product. For further questions call 1-866-781-1173 or contact your Acclarent Sales Representative.
Initiated
2012-01-30
Root cause
Component design/selection
Terminated
2013-08-29
Distribution
Nationwide Distribution (USA) only sixe 18 x 40 Worldwide distribution other sizes.Distirbution to France, Canada, Saudi Arabia
Recall number
Z-1095-2012
Recall status
Terminated
CDRH record ID
107235
FDA classified
2012-03-01
Recall event number
61070

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗