Z-1095-2017: Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to prov

Recall record Z-1095-2017 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1095-2017
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Jan 25, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.

Source facts

Action
Baxter sent an Urgent Product Recall on January 4, 2016, to all affected customers via U.S.P.S., first class mail. Customers were asked to: 1) Locate and remove all affected product lots from their facility. 2) Check if the Preventative Maintenance Kits were already installed in their Phoenix machines and if the case or if uncertain, contact their Baxter Service Representative at 1-800-525-2623 (Option 2) to schedule an appointment to remediate the machine. 3) Contact Baxter Health care Center for Service to arrange for return and credit. 4) Complete the enclosed Baxter Customer Reply Form and return it to Baxter by either fax or scanned email. All non-responding customers will be contacted via telephone. All affected product will be destroyed or discarded. Customers with questions were advised to call 800-437-5176.
Initiated
2017-01-04
Root cause
Mixed-up of materials/components
Terminated
2019-03-08
Distribution
Nationwide Distribution
Recall number
Z-1095-2017
Recall status
Terminated
CDRH record ID
152439
FDA record created
2017-01-25
Recall event number
76168

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗