Z-1096-06: Weck Hem-o-lok ML Polymer Ligating Clips; Rx, sterile, medium large size, non-absorbable polymer ligation

Recall record Z-1096-06 for Teleflex Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1096-06
Company in FDA record
Teleflex Medical
FDA classification date
Jun 9, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Hem-o-lok ligating clips may become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Hem-o-lok ligating clips are now contraindicated for use in ligating the renal artery during laparascopic nephrectomies in living donor patients.

Source facts

Action
Teleflex Medical sent Important Product Safety Information letters dated 4/18/06 to all of their customers who have purchased Hem-o-lok Ligating Clips from Teleflex. The letters informed the healthcare providers that Teleflex has been made aware of incidents in which the ligating clips (sizes L and XL) were reported to have become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Although none of the incidents appear to have involved any defect in or malfunction of the ligating clips, misapplication of the Hem-o-lok clips during such laparascopic procedures may not immediately be apparent and can have serious, even life-threatening consequences post-operatively. Therefore, Teleflex has added the following contraindication to the Instructions for Use accompanying the Hem-o-lok Polymer Ligating Clips: 'Contraindications': Hem-o-lok ligating clips are contraindicated for use in ligating the renal artery during laparoscopic nephrectomies in living donor patients. 'Caution': The clip must be latched to ensure proper ligation of the vessel or tissue. Inspect the ligation site after application to ensure proper closure of the clip. Security of the closure should be confirmed after ligation. The Hem-o-lok Polymer Ligating Clip is not designed for use as a tissue marker. Weck recommends that more than one clip be used to ligate the renal artery in procedures other than laparoscopic donor nephrectomy (see Contraindication, above). Application of more than one clip to all other vessels should be left to the surgeon's judgment.' Any questions concerning the letter were directed to Teleflex Medical at 800-334-9751, option #7, then option #5. Teleflex sent follow-up Important Product Correction Notice letters dated 8/2/06 to all of their Hem-o-lok Ligating Clips customers via registered mail on 8/7/06. The letters provide clarification of the 4/18/06 letter, reinforcing that the clips are contraindicated for use in ligating t
Initiated
2006-04-12
Root cause
Other
Terminated
2007-11-14
Distribution
Nationwide
Recall number
Z-1096-06
Recall status
Terminated
CDRH record ID
45799
FDA classified
2006-06-09
Recall event number
35316

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗