Open original FDA record ↗Recall record
- FDA identifier
- Z-1097-2016
- Company in FDA record
- Collagen Matrix Inc
- FDA classification date
- Mar 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time.
Source facts
- Action
- The firm, Collagen Matrix Inc., sent an email dated May 14, 2013 to its sole customer. The email described the product, problem and actions to be taken. The customer was instructed to return all product with expiration dates 12/2014 or earlier under Returned Goods, the firm will replace these products with new inventory; and return to Collagen Matrix, Inc., 15 Thornton Road, Oakland, NJ 07436 or quarantine products with expiration date 1/2016 until further notice. If you have any questions,call 201-405-1477 x 304 or email: www.collagenmatrix.com.
- Initiated
- 2013-05-13
- Root cause
- Process control
- Terminated
- 2017-06-21
- Distribution
- US Distribution to state of: MA.
- Recall number
- Z-1097-2016
- Recall status
- Terminated
- CDRH record ID
- 143566
- FDA classified
- 2016-03-11
- Recall event number
- 73206