Z-1098-2007: ARCHITECT AUSAB Controls list number 1L82-10 have two bottles with the following components: 1) Negative

Recall record Z-1098-2007 for Abbott Diagnostic International, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1098-2007
Company in FDA record
Abbott Diagnostic International, Ltd.
FDA classification date
Jul 24, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results and may impact the interpretation for a small percentage of results with grayzone or low-level reactive concentrations and higher proficiency sample results.

Source facts

Action
All ARCHITECT AUSAB U.S. customers will be contacted. New customers will be contacted through a Product Information Letter that will be included in all ARCHITECT AUSAB Reagent Kits until assay is restandardized. All ARCHITECT AUSAB U.S. customers will be contacted. The following communication types were used: Product Correction Letter Customer Reply Form > Product Information Letter to be included in new kits to be distributed. Letters were sent to customers by Federal Express Priority mail. The Customers were instructed to: Please follow their laboratory procedures for evaluation of these results Disregard the automated interpretation of results for low-level reactive samples. A customer reply form will accompany the Product Correction letter. Effectiveness Check Depth: 100% of customers receiving the Product Correction letter. The start date of communication was May 14,2007.
Initiated
2007-05-14
Root cause
Other
Terminated
2011-06-10
Distribution
Nationwide and Puerto Rico
Recall number
Z-1098-2007
Recall status
Terminated
CDRH record ID
52498
FDA classified
2007-07-24
Recall event number
37961

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗