Open original FDA record ↗Recall record
- FDA identifier
- Z-1098-2013
- Company in FDA record
- Carefusion Corporation
- FDA classification date
- Apr 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The recall was initiated because Carefusion has received reports of a communication error on the Alaris PC unit model 8015 with software version 9.12 when the Alaris EtCO2 module or Alaris SpO2 module is attached.
Source facts
- Action
- CareFusion sent an Urgent Medical Device Recall Notification letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and action to be taken. Customers are required to confirm receipt of the notification by returning the Recall Response card to Carefusion by postage-paid, self-addressed mail, fax, or email. Carefusion representatives will contact customers by telephone to schedule field remediation. For questions contact CareFusion Support Center at 888-562-6018, Customer Advocacy at 888-812-3266 and/or Technical Support at 888-812-3229.
- Initiated
- 2013-03-06
- Root cause
- Software design
- Terminated
- 2014-09-02
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia and China.
- Recall number
- Z-1098-2013
- Recall status
- Terminated
- CDRH record ID
- 116871
- FDA classified
- 2013-04-15
- Recall event number
- 64728