Open original FDA record ↗Recall record
- FDA identifier
- Z-1098-2016
- Company in FDA record
- Deerfield Imaging
- FDA classification date
- Mar 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Emergency helium venting lines (i.e. quench lines) may not have been properly installed.
Source facts
- Action
- Consignees were sent a IMRIS "Urgent Field Safety Notice" letter dated February 10, 2016. The letter described the problem and the product involved in the recall. The firm also stated that it is important that customers review the portion of the quench line vent described and ensure that it is installed to the specifications supplied as part of the Planning Guide. Customers were reminded that it is important to have an emergency plan in place in the event Helium gas escapes into the magnet room or any other area. For questions customers can contact their local customer support hotline or +1-866-475-0525.
- Initiated
- 2016-02-10
- Root cause
- Device Design
- Terminated
- 2017-04-05
- Distribution
- Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.
- Recall number
- Z-1098-2016
- Recall status
- Terminated
- CDRH record ID
- 144171
- FDA classified
- 2016-03-11
- Recall event number
- 73421