Open original FDA record ↗Recall record
- FDA identifier
- Z-1098-2017
- Company in FDA record
- Soft Computer Consultants, Inc.
- FDA record creation date
- Jan 26, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software error. Potential for incorrect results
Source facts
- Action
- Consignees were sent a recall notification, dated 10/06/2016, providing instructions for software correction. The firm indicates that the issue will be resolved with a mandatory hot fix. Hot fix will be released when its effectiveness is confirmed. Consignees are instructed to acknowledge receipt of the recall notification and grant permission for the recalling firm to load the hot fix when available. The firm can be contacted by calling Technical Support at 1-800-763-8522. Until the hot fix is available, consignees can still operate the software by exiting (Esc-Quit) the Test Verify option after each save changes (F12-Accept).
- Initiated
- 2016-10-06
- Root cause
- Software design (manufacturing process)
- Terminated
- 2019-08-21
- Distribution
- Worldwide Distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, WI, WV, and the countries of Hawaii and Puerto Rico Canada & Jamaica
- Recall number
- Z-1098-2017
- Recall status
- Terminated
- CDRH record ID
- 152049
- FDA record created
- 2017-01-26
- Recall event number
- 76046