Open original FDA record ↗Recall record
- FDA identifier
- Z-1099-2012
- Company in FDA record
- Clarity Medical Systems Inc
- FDA classification date
- Feb 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Still images extracted from video recordings may be mislabeled as to right and left eye.
Source facts
- Action
- Clarity Medical Systems, Inc. sent a Safety Alert letter dated February 7, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customer was instructed the following : 1. Do not use the video mode until the software has been updated. 2. Install the software update using the enclosed CD. The instructions for installation are also enclosed. This software update will correct the anomaly. 3. If you have extracted still images from video recordings, review all potentially affected patient files to determine if any right eye images are mislabeled as left eye images and/or left eye images are mislabeled as right eye images. If any images are mislabeled, make the appropriate corrections in the annotation section of the affected file. Consider whether there is any diagnostic impact or need to re-irnage the patient. We apologize for the inconvenience. If you have questions or concerns, please contact Clarity at 800-215-6005.
- Initiated
- 2012-02-02
- Root cause
- Software design
- Terminated
- 2012-06-07
- Distribution
- Worldwide Distribution -- USA (nationwide) including the states of TN, MO, TX, FL, NJ and Puerto Rico and countries of Italy, Portugal, Czech Republic, Thailand, Chile, Oman, Hong Kong, Russia, Philippines, Sweden, Finland, Ecuador, Australia and Canada.
- Recall number
- Z-1099-2012
- Recall status
- Terminated
- CDRH record ID
- 107470
- FDA classified
- 2012-02-24
- Recall event number
- 61180