Open original FDA record ↗Recall record
- FDA identifier
- Z-1099-2019
- Company in FDA record
- LivaNova USA Inc.
- FDA record creation date
- Apr 6, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
Source facts
- Action
- The recalling firm, LivaLova, contacted the first customer by phone on 9/14/2018 followed by an "IMPORT" letter, the first of which was dated 9/18/2018. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: keep these devices out of service until the investigation is complete. To return your device, please email 3T.US@livanova.com. If taking Serial Numbers 16S15335 and 16S15337 out of service will negatively impact your clinical practice, we can arrange for delivery of loaner devices during this investigation and,as necessary, while they undergo repeat Deep Cleaning Process. If you have any questions, contact V.P., Clinical, Quality, and Regulatory Services at 303-467-6306 or email: joan.ceasar@livanova.com.
- Initiated
- 2018-09-14
- Root cause
- Material/Component Contamination
- Terminated
- 2022-09-14
- Distribution
- US Distribution to states of: IN, MO, and OH.
- Recall number
- Z-1099-2019
- Recall status
- Terminated
- CDRH record ID
- 171458
- FDA record created
- 2019-04-06
- Recall event number
- 82365