Skip to content

Public FDA record

Z-1099-2019: SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung m

Recall record Z-1099-2019 for LivaNova USA Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1099-2019
Company in FDA record
LivaNova USA Inc.
FDA record creation date
Apr 6, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).

Source facts

Action
The recalling firm, LivaLova, contacted the first customer by phone on 9/14/2018 followed by an "IMPORT" letter, the first of which was dated 9/18/2018. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: keep these devices out of service until the investigation is complete. To return your device, please email 3T.US@livanova.com. If taking Serial Numbers 16S15335 and 16S15337 out of service will negatively impact your clinical practice, we can arrange for delivery of loaner devices during this investigation and,as necessary, while they undergo repeat Deep Cleaning Process. If you have any questions, contact V.P., Clinical, Quality, and Regulatory Services at 303-467-6306 or email: joan.ceasar@livanova.com.
Initiated
2018-09-14
Root cause
Material/Component Contamination
Terminated
2022-09-14
Distribution
US Distribution to states of: IN, MO, and OH.
Recall number
Z-1099-2019
Recall status
Terminated
CDRH record ID
171458
FDA record created
2019-04-06
Recall event number
82365

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗