Skip to content

Public FDA record

Z-1099-2021: Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in gre

Recall record Z-1099-2021 for Cytocell Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1099-2021
Company in FDA record
Cytocell Ltd.
FDA record creation date
Feb 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.

Source facts

Action
The firm began distributing recall notifications to customers on December 22, 2020. Customers were informed of the product issue. Customers were instructed to do the following: Immediately examine your inventory and quarantine all product subject to recall. The firm requests that you destroy the remaining inventory. The firm also suggests that laboratories review any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of any additional locus specific signal at 11q13. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Initiated
2020-12-22
Root cause
Under Investigation by firm
Terminated
2022-06-07
Distribution
Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.
Recall number
Z-1099-2021
Recall status
Terminated
CDRH record ID
185313
FDA record created
2021-02-08
Recall event number
87090

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗