Open original FDA record ↗Recall record
- FDA identifier
- Z-1100-2010
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Mar 11, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Toshiba America Medical Systems Inc is removing the PET-511BTM Multi-plane Transesophageal Transducer from the market after receiving several reports of injury regarding the transducer. Bending or buckling of the flexible portion of the PET-511 BTM multi-plane transesophageal transducer (TEE) may result in injury including perforation.
Source facts
- Action
- Toshiba America Medical Systems sent a Field Safety Notice to all customers on 5 MAY 2009 indicating its intent to remove the units from the market. The letter requested immediate cessation of use and return to the manufacturer site. It announced plans to replace unit with a new multi-plane transducer at no addle charge. The notice also included a description of the problem that occurred, the proper insertion technique into the esophageal passage, how to avoid and detect bending or buckling of the transducer, and what to do if the transducer does buckle. A revised safety manual and poster reflecting these precautions for the PET-51 IBTM was sent to each customer using this transducer. Shipments of the transducer from December 2008 also contained the Field Safety Notification materials. For additional information, contact the firm at (714) 669 2458.
- Initiated
- 2009-05-11
- Root cause
- Device Design
- Terminated
- 2012-03-22
- Distribution
- Nationwide
- Recall number
- Z-1100-2010
- Recall status
- Terminated
- CDRH record ID
- 86818
- FDA classified
- 2010-03-11
- Recall event number
- 53892