Open original FDA record ↗Recall record
- FDA identifier
- Z-1100-2012
- Company in FDA record
- Kerr Corporation
- FDA classification date
- Feb 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because Pentron Clinical has confirmed that the Breeze Self-Adhesive Resin Cement exhibits faster than expected gel and set times. If the work time is exceeded when using this material, it could possibly result in inadequate bond strength.
Source facts
- Action
- An "Urgent Medical Device Recall" letter was sent on 12/17/2010 via USPS 1st class mail to all their customers. The letter provides the customers with an explanation of the problem identified and the action to be taken. Customers were instructed to complete and fax back the enclosed recall return form. Customers with questions were instructed to call (800) 551-0283.
- Initiated
- 2010-12-02
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-02-24
- Distribution
- Worldwide Distribution - Nationwide (USA) and the countries of Vietnam, Jordan,Turkey, and Italy.
- Recall number
- Z-1100-2012
- Recall status
- Terminated
- CDRH record ID
- 107626
- FDA classified
- 2012-02-24
- Recall event number
- 61235