Skip to content

Public FDA record

Z-1100-2017: AutoDELFIA Plate Processor , Fluorometer, for clinical use Product Number: 1235-5110 Product Usage: Th

Recall record Z-1100-2017 for PerkinElmer Health Sciences, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1100-2017
Company in FDA record
PerkinElmer Health Sciences, Inc.
FDA record creation date
Oct 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

AutoDELFIA System with Version 3.0 Software: Potential plate loading failures

Source facts

Action
PerkinElmer issued letter on 12/21/16 to customers with AutoDELFIA Systems running version 3.0 Manager Software (revisons 1 and 2). The issue cannot be corrected by the customer therefore it is recommended the AutoDELFIA System be operated in the default English language variant until the AutoDELFIA Manager Software can be updated. The AutoDELFIA System with version 3.0 Manager Software performs as intended when the selected language variant is English. Customers are recommended to implement a software update to be deployed by PerkinElmer to resolve the issue. AutoDELFIA Manager Software version 3.0 revision 1 or 2 will be updated through the correction to revision 3 or the latest released revision
Initiated
2016-12-21
Root cause
Software design
Terminated
2018-10-03
Distribution
Worldwide Distribution - US Nationwide in the state of MN and countries of China, India, Philippines, France, South Korea, Vietnam, Taiwan, Morocco, Kazachstan, UK, Spain, and Israel,
Recall number
Z-1100-2017
Recall status
Terminated
CDRH record ID
151811
FDA record created
2018-10-11
Recall event number
75949

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗