Z-1100-2021: Arjo Flat Dynamic Positioning System (DPS) - Product Usage: used in conjunction with passive clip slings

Recall record Z-1100-2021 for Arjohuntleigh Magog, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1100-2021
Company in FDA record
Arjohuntleigh Magog, Inc.
FDA record creation date
Feb 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Risk of Sling Clip breakage while in use with obsolete Flat Dynamic Positioning System (DPS)

Source facts

Action
On 10/06/2020, the firm sent a letter to its consignees with the following instructions. Arjo has become aware that there is a likelihood that a sling clip may break (Image 2) when used with the models of spreader bars referenced above that are out of their lifespan (see note 1 below). The continued use of the obsoleted DPS can lead to complete clip breakage and subsequently to its detachment from the DPS attachment point. As a consequence, the resident may fall off and potentially sustain serious injuries as a consequence. Arjo is taking this matter very seriously and recommends replacement of the obsoleted DPS on ceiling lift or floor lift systems. Note 1: The probability of the event may increase if the DPS is used with a Clip Sling that has exceeded its expected lifespan of 2 years or more than 1.5 years for Passive Clip Slings models MAA2080M, MAA2050M, MAA2040M, MAA2070M and MLAAS2000. Next Steps 1. Verify whether you possess DPS within the range of this Notice and ensure that all caregivers and users of Arjo ceiling lift or floor lifts are made aware of this Field Safety Notice (FSN). 2. Complete and sign the enclosed Customer Response Form and return this document to your local Arjo office. Contact your local Arjo office to take the appropriate decision regarding your obsolete device(s) and select the best solution for your needs. NOTE: Until you find with our Arjo representative the most appropriate solution for your patient transfer needs, the device may stay in use strictly if those two conditions are fulfilled; A. Before each patient transfer, ensure to follow the recommendation of the Sling Instruction for Use and check all parts of the sling are free of any damage B. Ensure the sling used is within its intended lifetime (2 years or 1.5 years for Passive Clip Slings models MAA2080M, MAA2050M, MAA2040M, MAA2070M and MLAAS2000). In addition, if the label of the sling is faded, unreadable or damaged, ensure sling disposal as recomme
Initiated
2020-10-06
Root cause
Nonconforming Material/Component
Terminated
2025-07-23
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, GA, IA, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, MT, NC, NE, NH, NY, OH, OK, SC, SD, TN, TX, UT, VA, WI and the countries of Australia, Belgium, Canada, Czech Republic, France, Italy and Poland.
Recall number
Z-1100-2021
Recall status
Terminated
CDRH record ID
184078
FDA record created
2021-02-09
Recall event number
86578

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗