Skip to content

Public FDA record

Z-1100-2025: The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrica

Recall record Z-1100-2025 for Ithera Medical Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1100-2025
Company in FDA record
Ithera Medical Gmbh
FDA classification date
Feb 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).

Source facts

Action
The firm sent an "Urgent Field Safety Notice" letter dated 12/19/2024 to its consignees. FSN 3746 has been distributed to all customers to correct a wrong value in the user manual for the NOHD. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Take note of amendment/reinforcement of Instructions For Use (IFU) 1. Confirm the receipt of this FSN 2. Forward this information and the updated IFU to the person responsible for laser safety at your location (such as Laser Safety Officer) 3. The person responsible for laser safety at your location performs a new assessment of the laser safety consideration and to identify any additionally required work safety related measures 4. Training of all users of the MSOT Acuity Echo device on any additionally required work safety related measures, if applicable. The new calculation of the NOHD resulting in a higher value due to the use of a different calculation method. The user manual has been updated accordingly, and users have been informed. iThera will provide an updated user manual (IM_2.03_EN_III) with the corrected value for NOHD. If you have any questions, call 203-997-5110 US, +49-80-700749; EXT21 OUS or email: josef.konardl@ithera-medical.com.
Initiated
2025-01-08
Root cause
Radiation Control for Health and Safety Act
Distribution
Worldwide distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Germany, Italy, Netherlands, Portugal, Switzerland, and United Kingdom.
Recall number
Z-1100-2025
Recall status
Open, Classified
CDRH record ID
212392
FDA classified
2025-02-07
Recall event number
96234

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗