Skip to content

Public FDA record

Z-1101-2021: GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: In

Recall record Z-1101-2021 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1101-2021
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Feb 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.

Source facts

Action
The firm issued an urgent medical device correction notice by letter on 12/10/2020. The letter states: "You can continue to use your Precision 600 FP monitor suspension under the following condition. If you observe any abnormal movement or misalignment of the monitor suspension in day to day performance controls, discontinue use and contact your GE Healthcare Service Representative immediately." Customers with questions or concerns may contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2020-12-10
Root cause
Other
Terminated
2022-11-18
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV
Recall number
Z-1101-2021
Recall status
Terminated
CDRH record ID
185143
FDA record created
2021-02-09
Recall event number
87061

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗