Z-1101-2023: RipCord Syndesmosis Button, REF: STA001K

Recall record Z-1101-2023 for TriMed Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1101-2023
Company in FDA record
TriMed Inc.
FDA classification date
Feb 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.

Source facts

Action
On 12/22/22, Safety Alert/Advisory Notices were emailed to consignees who were told the following: The updated surgical technique can be found at trimedortho.com/surgical-techniques/ This notice needs to be shared with those who need to be aware within your organization. Acknowledgment and Receipt forms were sent with the notices to be filled out and returned to the firm via the email address below. 1. Deliver the suture passer and pull suture (white/blue) through the soft tissue on the medial side. 2. Maintain gentle tension on the sutures while passing the medial button through the tunnel. Wrapping the pull suture around a finger of one hand and holding the device sutures on the opposite side with the other hand helps maintain gentle, continuous tension as the button is passed and maintains alignment of the medial button with the axis of the tunnel. 3. Gently slide the sutures back and forth to ensure that there are no obstructions and that the sutures are aligned with the bone tunnel. 4. Continue to gently oscillate as the medial button starts to engage the hole and proceed with a swift, consistent pull to deliver the medial button out the medial side. Avoid excessive movement of the fibula relative to the tibia which can misalign the bone tunnels. If excessive resistance is felt, withdraw the button and try again. 5. Once the medial button clears the medial cortex pull back on the device suture to seat the button; cut and remove the pull suture. If you have any questions, please contact quality@trimedortho.com On 2/14/23, recall notices were sent to customers who were asked to do the following: 1) Stop further use of the device and return any unused products to the recalling firm. 2) Provide a copy of this recall notice to the hospital/surgeon/physician that is using the device. 3) Complete and return the response form.
Initiated
2022-12-22
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of IL, PA, TX, GA, TN, AL, MO, FL, AR, LA, HI, OH, TX, KS
Recall number
Z-1101-2023
Recall status
Open, Classified
CDRH record ID
197884
FDA classified
2023-02-07
Recall event number
91478

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗