Skip to content

Public FDA record

Z-1101-2025: . Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 T

Recall record Z-1101-2025 for Nobel Biocare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1101-2025
Company in FDA record
Nobel Biocare
FDA classification date
Feb 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Source facts

Action
On 01/10/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Nobel Biocare became aware through internal quality control one of the plastic components of the packaging that is used to hold the OsseoShaper in place may contain a manufacturing error. This results in a sharp edge that could puncture the blister packaging, thereby breaching the sterile barrier. Customers and Distributors are instructed to: 1. Inspect their stock and quarantine affected devices. The use of affected devices should be ceased immediately. 2. For Distributors, complete attached Distributor Acknowledgement Form with information collected from end user and return the firm to Nobel Biocare via email to us.cs.complaints@nobelbiocare.com with in 5 days of receipt of this notice. 2. Complete attached Customer Acknowledgment Form, even if customers do not have any affected stock, and return it to Nobel Biocare, via email to us.cs.complaints@nobelbiocare.com, within 5 days of receipt of this notice. 3. Return all affected stock on hand to Nobel Biocare using the shipping label attached to this notice. 4. Ensure relevant staff members are informed of this recall. If products were supplied or transferred to another facility or organization, inform these facilities of the recall by providing a copy of this notice. For question or further support, contact local customer support representative at 800-322-5001 ext. 1527 or at us.cs.complaints@nobelbiocare.com
Initiated
2025-01-10
Root cause
Process control
Distribution
U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.
Recall number
Z-1101-2025
Recall status
Open, Classified
CDRH record ID
211856
FDA classified
2025-02-05
Recall event number
96047

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗