Skip to content

Public FDA record

Z-1102-2007: Cell Search Circulating Tumor Cell Kit (Epithelial) Veridex LLC, a Johnson & Johnson Company IVD Veridex,

Recall record Z-1102-2007 for Veridex, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1102-2007
Company in FDA record
Veridex, LLC
FDA classification date
Jul 24, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An increased incidence of aggregates was observed in Lot 0036 and 0036A. The presence of aggregates will cause customers to invalidate patient results.

Source facts

Action
All Direct customers and customers of distributors received a customer notification informing them of the increased incidence of aggregates in cartridges on April 30, 2007 by Fed Ex overnight. This letter instructed them to discontinue use and to discard any remaining inventory. This was followed by a notification on May 24, 2007 by Fed Ex overnight to all US Distributors which also requested that they immediately stop using and/or distributing kits and to destroy any remaining inventory. Distributors were also asked to notifiy customers who received product. All foreign affiliate consignees were also informed on this date by e-mail and instructed to notify their customers.
Initiated
2007-04-17
Root cause
Other
Terminated
2007-12-06
Distribution
The kits were distributed to hospitals and laboratories nationwide. They were also sent to 7 J&J affiliated companies in France, Japan, Canada, England, Germany, Italy and Spain.
Recall number
Z-1102-2007
Recall status
Terminated
CDRH record ID
53161
FDA classified
2007-07-24
Recall event number
38206

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗