- FDA identifier
- Z-1102-2008
- Company in FDA record
- Seradyn, Inc.
- FDA classification date
- Apr 29, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
False negative test results due to Interfering substances (heterophile antibodies) in the patient's blood may produce erroneously low results.
Source facts
- Action
- Consignees were notified of the problem via Urgent Medical Device Correction letter dated 12/20/07. The letter provided information on the issues concerning the recall. The firm will revise the package insert to address the potential occurrence of false negative bias. If inconsistencies with Vancomycin level exist, confirm with an alternative test method. For further information please contact 1-800-428-4072.
- Initiated
- 2007-12-20
- Root cause
- Device Design
- Terminated
- 2008-12-30
- Distribution
- Worldwide Distribution --- USA including states of California and New York, and countries of Australia, Belgium, Canada, Finland, Germany, Ireland, Japan, Spain and United Kingdom.
- Recall number
- Z-1102-2008
- Recall status
- Terminated
- CDRH record ID
- 66885
- FDA classified
- 2008-04-29
- Recall event number
- 46150