Open original FDA record ↗Recall record
- FDA identifier
- Z-1102-2017
- Company in FDA record
- PerkinElmer Health Sciences, Inc.
- FDA record creation date
- Oct 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
AutoDELFIA System with Version 3.0 Software: Potential plate loading failures
Source facts
- Action
- PerkinElmer issued letter on 12/21/16 to customers with AutoDELFIA Systems running version 3.0 Manager Software (revisons 1 and 2). The issue cannot be corrected by the customer therefore it is recommended the AutoDELFIA System be operated in the default English language variant until the AutoDELFIA Manager Software can be updated. The AutoDELFIA System with version 3.0 Manager Software performs as intended when the selected language variant is English. Customers are recommended to implement a software update to be deployed by PerkinElmer to resolve the issue. AutoDELFIA Manager Software version 3.0 revision 1 or 2 will be updated through the correction to revision 3 or the latest released revision
- Initiated
- 2016-12-21
- Root cause
- Software design
- Terminated
- 2018-10-03
- Distribution
- Worldwide Distribution - US Nationwide in the state of MN and countries of China, India, Philippines, France, South Korea, Vietnam, Taiwan, Morocco, Kazachstan, UK, Spain, and Israel,
- Recall number
- Z-1102-2017
- Recall status
- Terminated
- CDRH record ID
- 151819
- FDA record created
- 2018-10-11
- Recall event number
- 75949