Skip to content

Public FDA record

Z-1102-2025: LIFEPAK 35 ECG cable REF 11111-000041

Recall record Z-1102-2025 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1102-2025
Company in FDA record
Physio-Control, Inc.
FDA classification date
Feb 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions

Source facts

Action
On 01/21/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that Stryker discovered that an incorrect version of the Instructions for Use (IFU) for the 3-wire extended precordial ECG cable for the LIFEPAK 35 were distributed to customers. This version did not include information regarding disinfection of the 3-wire cable. Customer are instructed to: 1. Immediately check internal inventory to locate the product listed on the attached business reply form. 2. Remove and quarantine any affected product from use. 3. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, email the enclosed Business Reply Form (BRF) by January 31st, 2025 date, to RSRecall@Stryker.com. 4. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for the return and replacement of their product(s). 5. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until replacement. 6. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: rsdevicetracking@stryker.com b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ If you have any questions or concerns, contact Customer Service +1 800 442 1142 or by email RSRecall@Stryker.com.
Initiated
2025-01-21
Root cause
Labeling Change Control
Distribution
U.S. Nationwide distribution in the states of CA, FL, GA, IA, IN, KY, MD, MI, MO, NE, NV, NY, OH, OR, SC, TN, TX, and WI.
Recall number
Z-1102-2025
Recall status
Open, Classified
CDRH record ID
211857
FDA classified
2025-02-05
Recall event number
96048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗