Skip to content

Public FDA record

Z-1103-2014: Optiflux F200NR High Flux Dialyzers; Product Number: 0500320N. Single Use acute or chronic hemanalysis t

Recall record Z-1103-2014 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1103-2014
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Feb 26, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dialysate Port Leak During Priming of the Dialyzers

Source facts

Action
Fresenius Medical issued Notice to Clinics and Clinics with Patients Letters Dated - 1st Mailing - February 4, 2014 and Patients Letters Dated - February 5, 2014 - these letters contained the correct information.Corrected letters to Clinics and Clinics with Patients issued 2/14/14 via certified mail with return receipt and fax-back form . Users are advised the leak may be stopped by tightening the dialysate port cap more securely. Questions associated with this product notification to your Customer Service Team at 1-800-323-5188.
Initiated
2014-02-04
Root cause
Under Investigation by firm
Terminated
2021-05-18
Distribution
Worldwide distribution: USA (nationwide) and the country of: Mexico.
Recall number
Z-1103-2014
Recall status
Terminated
CDRH record ID
125235
FDA classified
2014-02-26
Recall event number
67313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗