Open original FDA record ↗Recall record
- FDA identifier
- Z-1103-2020
- Company in FDA record
- Fridababy LLC
- FDA classification date
- Jan 30, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
FridaBaby has been made aware through customer reports that the product is being used outside of the labeled directions. Specifically, repeated chewing or biting on the SmileFrida, the ToothHugger has caused deterioration in the outer layer of the brush and in very rare occasions, may cause the hard plastic brush head to break, exposing a small metal piece welded inside the brush head.
Source facts
- Action
- Consignees will be notified via email with recall letter. The letter identifies affected product, stated reason for recall, and request to inspect the product for evidence of misuse and discard if necessary. If product has not been used, the consignee is reminded to follow the product labeling. A response to the recall notification via email is requested.
- Initiated
- 2020-01-27
- Root cause
- Use error
- Distribution
- Nationwide. International distribution to Canada, Mexico, Nicaragua, Dominican Republic, Honduras, Chile, Peru, Qatar. No US governmental consignees.
- Recall number
- Z-1103-2020
- Recall status
- Open, Classified
- CDRH record ID
- 179086
- FDA classified
- 2020-01-30
- Recall event number
- 84707