Z-1103-2021: Universal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LP

Recall record Z-1103-2021 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1103-2021
Company in FDA record
DePuy Orthopaedics, Inc.
FDA record creation date
Feb 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots.

Source facts

Action
On 01/12/2021, DePuy Synthes issued an Urgent Medical Device Recall (removal) via letter due to taper dimensions of certain Universal Femoral Sleeves may be out of specification. The Universal Femoral Sleeves are used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products. DePuy is recommending that customers return the recalled devices.
Initiated
2021-01-08
Root cause
Nonconforming Material/Component
Terminated
2024-01-17
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, FL, IN, MI, MN, MS, NC, NY, OH, PA, TX, VA and the countries of Canada, China, Germany, Puerto Rico, Switzerland.
Recall number
Z-1103-2021
Recall status
Terminated
CDRH record ID
185512
FDA record created
2021-02-09
Recall event number
87189

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗