Open original FDA record ↗Recall record
- FDA identifier
- Z-1104-06
- Company in FDA record
- Hitachi Medical Systems America Inc
- FDA classification date
- Jun 10, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.
Source facts
- Action
- The firm''s Field Service personnel will apprise the customers of this situation during scheduled visits at each customer location. During this visit, the corrected software will be installed by the Hitachi Field Service Technician. The installation of the software upgrades by the service technicians began on 4/28/2006. The firm anticipates that the correction of all units in their inventory as well as those units installed out-in-the-field will becompleted by 6/15/2006.
- Initiated
- 2006-04-28
- Root cause
- Other
- Terminated
- 2008-11-24
- Distribution
- The devices were distributed to five (5) customers located in four (4)states: Colorado, Ohio, Nebraska and Texas
- Recall number
- Z-1104-06
- Recall status
- Terminated
- CDRH record ID
- 45322
- FDA classified
- 2006-06-10
- Recall event number
- 35100