Software anomaly. A software defect in the referenced systems can cause a slice indicator (reference line) to be mis-positioned on the cross-reference image when it is magnified. This image error could result in a possible patient mis-diagnosis.
Source facts
Action
On 3/17/2006, the firm began conducting a survey of all afftected customer sites in order to determine if corrected software has been installed on the unit(s) at each customer location. If the corrected software has not been installed, the firm''s Field Service Engineers will install software upgrades as necessary.
Initiated
2006-03-17
Root cause
Other
Terminated
2008-11-13
Distribution
Nationwide-The devices were distributed to consignees located througout the United States.
Recall number
Z-1105-06
Recall status
Terminated
CDRH record ID
45324
FDA classified
2006-06-10
Recall event number
35103