Skip to content

Public FDA record

Z-1105-2018: smith&nephew, LEGION HK FEMORAL WEDGE (WEDGE AND SCREW), 5 MM, SIZE 5, REF 71422159, QTY: (1), STERILE

Recall record Z-1105-2018 for Smith & Nephew, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1105-2018
Company in FDA record
Smith & Nephew, Inc.
FDA record creation date
Mar 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The screws package with devices within lot 14AAP6352A were 5 MM instead of 10MM and are too short to be used with the Legion HK Distal Wedge Size 5.

Source facts

Action
The firm initiated their recall by email on February 1, 2018. The letter stated the following: "Required Actions: 1. Please inspect your inventory and locate any unused devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the batch numbers and quantities of each batch that you are returning in the appropriate boxes below. 4. Complete the remainders of the form sign and send to FieldActions@smith-nephew.com or fax to 901-566- 7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com." For further questions, please call (901) 396-2121.
Initiated
2018-02-01
Root cause
Process control
Terminated
2018-10-22
Distribution
Worldwide Distribution - US Distribution to the states of: NY, IL, NJ, TX, CO, and to the countries of : Belgium, Great Britain, Australia, Germany, Netherlands, Norway.
Recall number
Z-1105-2018
Recall status
Terminated
CDRH record ID
161747
FDA record created
2018-03-19
Recall event number
79247

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.