Open original FDA record ↗Recall record
- FDA identifier
- Z-1106-06
- Company in FDA record
- MRL, Inc., A Welch Allyn Company
- FDA classification date
- Jun 22, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Welch Allyn AED 20 Defibrillators may display a 'Defib Comm' error message on the device display during use which may result in a terminal failure of the device to analyze the patient's ECG and deliver the appropriate therapy.
Source facts
- Action
- MRL issued a press release on 6/15/06, and began sending Urgent Medical Device Recall letters on the same date to the direct accounts informing them of the possibility of the defibrillator going into a terminal failure due to a U25 IC socket susceptible to defibrillation circuit board communication errors. The accounts were asked to respond to the notification witihin 5 working days by faxing or mailing the enclosed response form, or calling Welch Allyn-MRL Technical Support at 1-800-462-0777. A loaner AED 20 will then be sent to the consignee with instrucitons on how to return their unit for correction.
- Initiated
- 2006-06-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2007-07-19
- Distribution
- Worldwide-including USA, Australia, Canada, China, France, Germany, Greece, Ireland, Israel, Italy, Latvia, Nicaragua, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, United Arab Emirates and the United Kingdom.
- Recall number
- Z-1106-06
- Recall status
- Terminated
- CDRH record ID
- 46430
- FDA classified
- 2006-06-22
- Recall event number
- 35587