Therapy cable doesn't meet firm's standards/may not maintain connections sufficient to support ECG monitoring, external pacing or defibrillation; ECG connector with higher than normal connection resistance resulting in intermittent ECG signal; therapy connector has potential for impact damage.
Source facts
Action
The firm reviewed records and/or contacted customers by telephone to verify affected devices used in the out of hospital mobile environment. All affected customers were notified by a letter dated May 2004. The firm''s service representative will contact customers to inspect and correct the devices.
Initiated
2004-05-05
Root cause
Other
Terminated
2006-05-06
Distribution
The firm distributes to distributors, hospitals, EMT-Ds and paramedics nationwide. The firm also distibutes the product internationally.
Recall number
Z-1107-04
Recall status
Terminated
CDRH record ID
32993
FDA classified
2004-07-20
Recall event number
29025