Z-1107-2021: EliA dsDNA Well - Product Usage: intended for the In-Vitro quantitative measurement of IgG antibodies dir

Recall record Z-1107-2021 for Phadia US Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1107-2021
Company in FDA record
Phadia US Inc
FDA classification date
Feb 11, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).

Source facts

Action
Phadia US Inc. (Initial Importer/Distributor) issued notification to consignees via phone and FedEx overnight delivery Dec. 17-18 . Letter states reason for recall, health risk and action to take: Customers are asked to perform the following steps: 1.Scrap EliA dsDNA Well lot BFA3Y/0142 and order a replacement free of charge 2. Assess test results from EliA dsDNA Well lot BFA3Y/0142 and determine if retesting of samples is warranted in accordance with internal operating procedures. 3. Complete the Acknowledgement Form in the Medical Device Field Safety Notice and return the response to the contact person described. Any questions concerning this issue, contact Phadia US Technical Support at 1-800-346-4364, option #2.
Initiated
2020-12-14
Root cause
Under Investigation by firm
Terminated
2022-08-04
Distribution
US Nationwide distribution.
Recall number
Z-1107-2021
Recall status
Terminated
CDRH record ID
185460
FDA classified
2021-02-11
Recall event number
87161

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗