Z-1107-2023: TrackX Insight Base Modular Instrument Tracker, 5L, REF #604-3TX, Sterile; and TrackX Insight Base Modul

Recall record Z-1107-2023 for Trackx Technology Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1107-2023
Company in FDA record
Trackx Technology Inc
FDA classification date
Feb 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device is breaking and falling off the Nuvasive T-Handle with Hudson Adapter on certain Nuvasive XLIF instruments during surgery when malleted.

Source facts

Action
The recall was initiated via phone call on 10/24/2022 and 10/27/2022 (as well as subsequent in-person meetings) to two of the physicians who use the product in the affected application. The physicians were notified that the products can fragment when attached to a T-Handle and Nuvasive XLIF disectomy instruments while malleting and TrackX is limiting the use of those two bases to situations in the Nuvasive XLIF and Lateral Interbody procedures that do not require the malleting of the T-handle. The product is to be used in a limited manner and not on the Nuvasive T-Handle with Nuvasive XLIF Disectomy instruments. Letters dated 12/30/2022 began issuing via hand delivery or via DocuSign to all customers who have received the products. The letter informed the customer of the issue and said the IFU was revised with the application change and warning. The customer was informed that if they have further distributed or transferred the product to another location, they are to notify the recipients of the field action. A distribution list was provided to show the product that was shipped to the customer. An Acknowledgement and Receipt Form was enclosed along with the revised IFU (Warnings and Precautions section).
Initiated
2022-10-24
Root cause
Device Design
Terminated
2026-09-10
Distribution
US Nationwide distribution in the states of AZ, FL, LA, NC, PA, VA, WA, WI.
Recall number
Z-1107-2023
Recall status
Terminated
CDRH record ID
197510
FDA classified
2023-02-09
Recall event number
91338

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗