Z-1108-2017: Spare Disk Remover Module; Product Number: 61001820 Intended for in vitro quantitative and qualitative d

Recall record Z-1108-2017 for PerkinElmer Life and Analytical Sciences, Wallac, OY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1108-2017
Company in FDA record
PerkinElmer Life and Analytical Sciences, Wallac, OY
FDA classification date
Jan 27, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.

Source facts

Action
The recall was initiated on January 5, 2017 via email sent through the PerkinElmer Technical Support group located in Akron Ohio. Actions to be taken by user: 1. As instructed in the analyte specific kit inserts, the use of Kit Controls is required to assure the day-to-day validity of results. Patient results should only be reported if control results for the assay meet the laboratorys established criteria for acceptability. 2. If a movement error occurs customers will be directed to contact their local PerkinElmer field service representative. The movement error may indicate deterioration of the axle and therefore the use of the instrument should be avoided until field service personnel have inspected the status of the ball bushing axles or replaced affected ball bushing axles. Actions to be taken by the PerkinElmer: As a corrective action, PerkinElmer will replace the affected ball bushing axles in all the GSP instruments identified in this FSCA/Recall. The replacement may take place either as part of the initial installation activity before the instrument is used, scheduled periodic maintenance, or during a dedicated service visit. This field correction is considered as mandatory by the manufacturer. Customers will be contacted by PerkinElmer to arrange the field service visit.
Initiated
2017-01-05
Root cause
Component change control
Terminated
2017-08-24
Distribution
MA, SC, TX, AL OUS: China, Mexico, Peru, Morocco, Italy, Finland, Ecuador, Brazil
Recall number
Z-1108-2017
Recall status
Terminated
CDRH record ID
152448
FDA classified
2017-01-27
Recall event number
76165

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗