Z-1108-2021: Cascade Abrazo aPTT Test Card - Product Usage: are to be used with the Cascade Abrazo analyzer and are in

Recall record Z-1108-2021 for Helena Laboratories, Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1108-2021
Company in FDA record
Helena Laboratories, Corp.
FDA record creation date
Feb 11, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Intermittent potential for shorter than expected clot times.

Source facts

Action
On 01/05/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via email to it distributors informing them that this Recall has been identified due to the potential for shorter than expected clot times on abnormal patients and on abnormal controls on some cards from specific lots and the use of these products may include difficulty achieving abnormal control results in range and incorrectly low patient results. Distributors were instructed to: 1) Immediately examine their inventory and quarantine the specific product lots 2) If they have further distributed the affected product, they are to identify these customers and notify them at once. In addition, the notification to their customers may be enhanced by including a copy of the recall notification letter, which must include instruction of what customers should do with the recalled product. 3) Complete and return enclosed response form as soon as possible Any questions, email or contact Dr. Jessica Jones Hanka at jhanka@helena.com or 1-800-231-5663 ext. 1177. Distributors were contacted by email and requested to evaluate inventory while the Recalling Firm investigated issue . Distributors were formally notified of recall on 1/5/2021 and provided with a template for a notification letter to translate and send to end users in each country. End users will be asked to return unused product to the distributor, and distributors will forward all returned product to Helena Labs. Distributors will be asked to provide an account of all product distributed. Effectiveness will be determined by evaluating the total product returned and the total product accounted for based on information provided by distributors.
Initiated
2021-01-05
Root cause
Under Investigation by firm
Terminated
2024-11-19
Distribution
Worldwide distribution - U.S. Nationwide distribution and the countries of Germany, Italy, Romania, Australia, Chile.
Recall number
Z-1108-2021
Recall status
Terminated
CDRH record ID
185445
FDA record created
2021-02-11
Recall event number
87151

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗