Z-1110-2019: AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System.

Recall record Z-1110-2019 for PROCEPT BIOROBOTICS CORPORATION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1110-2019
Company in FDA record
PROCEPT BIOROBOTICS CORPORATION
FDA record creation date
Apr 8, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.

Source facts

Action
On 03/06/19, the firm, Procept BioRobotics, sent an"URGENT: VOLUNTARY MEDICAL DEVICE RECALL" letter via priority overnight mail with delivery confirmation to customers with affected devices. The letter the described the product, problem and actions to be taken. Customers were instructed to discontinue use and segregate affected devices. In addition, customers were asked to pass along the recall notice to all those who need to be aware of it within their organization and to complete and return the Recall Acknowledgement Form via Email: cs@procept-biorobotics.com or Fax to: +1.888.285.3777. Once the firm receives acknowledgement forms listing affected product, the firm will provide instructions on how to return affected product. Customers with questions are encouraged to call customer service at 1-650-232-7222. On 03/06/19, Courtesy Notification: Voluntary Medical Device Recall notices were mailed via priority overnight mail with delivery confirmation to customers who received non-affected devices. Customers who did not receive affected product were advised that no actions needed to be taken. Customers with questions are encouraged to call customer service at 1-650-232-7222. If you have any questions, please feel free to contact your local Aquablation Consultant or PROCEPT Customer Service by telephone at +1.650.232.7222 or email cs@procept-biorobotics.com.
Initiated
2019-03-06
Root cause
Process control
Terminated
2020-11-13
Distribution
Worldwide Distribution: US (nationwide) state to: TX, FL, IN, CA, IL, VA, MI, PA.; and (OUS) countries of: Italy, Germany, France, Poland, United Kingdom, Spain, Canada and Australia.
Recall number
Z-1110-2019
Recall status
Terminated
CDRH record ID
171653
FDA record created
2019-04-08
Recall event number
82462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗