Open original FDA record ↗Recall record
- FDA identifier
- Z-1111-2014
- Company in FDA record
- Roche Molecular Systems, Inc.
- FDA classification date
- Feb 26, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Numerous complaints have been filed that leaking cobas PCR media 4.3 mL IVD have been received from different lots of the cobas PCR Urine kits, cobas PCR Female Swab kits, and cobas PCR media kits.
Source facts
- Action
- Roche Molecular Diagnostics Inc. sent an Urgent Medical Device Correction letter/Fax-back form dated January 6, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the test-skpecific Instruction For Use for handling of cobas PCR media 4.3 ml IVD. The use of protective equipment should be worn as indicated in the labeling. Inspection each tube and/or blister packet for evidence of leakage prior to use. If leakage of the cobas PCR media 4.3 mL IVD is observed, do not use the tubes. Discard the affected tubes per local guidelines and contact Roche Support Network Customer Support Center at 1-800-526-1247. Complete the attached fax form and fax it to 1-855-695-8564. File this Urgent Medical Device Correction (UMDC) for future reference. Customers with questions were instructed to call 1-800-526-1247. For questions regarding this recall call 908-253-7200.
- Initiated
- 2014-01-06
- Root cause
- Device Design
- Terminated
- 2014-08-07
- Distribution
- Nationwide Distribution
- Recall number
- Z-1111-2014
- Recall status
- Terminated
- CDRH record ID
- 125455
- FDA classified
- 2014-02-26
- Recall event number
- 67441