Skip to content

Public FDA record

Z-1111-2025: Temporary Titanium Abutments

Recall record Z-1111-2025 for DDS Lab. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1111-2025
Company in FDA record
DDS Lab
FDA classification date
Feb 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain definitive dental implant restoration cases may have included a temporary titanium abutment (cylinder), rather than a definitive titanium abutment, which may lead the titanium abutment fracture, which may risk loosening of the affixed restoration; fragments could also injure oral mucosa, and could be aspirated.

Source facts

Action
On 12/6/2024, correction notices were mailed to customers asking them to do the following: Patients that received a dental implant restoration processed with a temporary titanium abutment (cylinder) rather than a definitive titanium abutment should be contacted to inform them about this Correction. Dental providers should continue to monitor a restoration subject to this Correction for potential problems, consistent with their clinical standard of care for all dental implant restorations. Firm is actively collaborating with dental support organizations and dental offices to ensure patients are notified of this event. If a dental provider identifies a potential problem with a restoration subject to this Correction, please contact the firm. To use case ID to access additional information including patient names, use the following steps in the My DDS Lab Portal: 1) Access the MyDDSLab Portal (https://my.ddslab.com/Login.aspx) 2) Can't remember your office's log-in or password? Call 877-966-9931 to request a reset. 3) Once logged into the portal, navigate to the "Case Management" tab. 4) Search by Case ID to find the patient's name and additional details about the fabrication. 5) Once in the case, you can also choose to download the invoice that will have more specific information about the fabrication. Please direct any inquiries to the vendor supporting the market correction effort via phone at 855-544-4821 Monday thru Friday 8am to 7pm ET or by email at ddslab6193@sedgwick.com.
Initiated
2024-12-06
Root cause
Component design/selection
Distribution
US Nationwide distribution including in the states of FL, TN, TX, MA, NH, CT, ME, WA, IN, CA, PA, NV, MO, IL, AK, MN, VT, OH, MI, VA, WI, AL, AZ, KY, MT, OK, CO, LA, GA, NY, WV, NC, NJ, MD, OR, DE, SC, ID, KS, IA, AR, NM, HI, DC, MS, NE, RI, UT.
Recall number
Z-1111-2025
Recall status
Open, Classified
CDRH record ID
211900
FDA classified
2025-02-05
Recall event number
96064

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗