Z-1112-2017: SwishTapered Implant

Recall record Z-1112-2017 for Implant Direct Sybron Manufacturing, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1112-2017
Company in FDA record
Implant Direct Sybron Manufacturing, LLC
FDA classification date
Jan 28, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The extender product code (9037-03) that is packaged with the SwishTapered (Part Number 933308; Lot Number 73241) Implant was packaged in the incorrect position. This discrepancy may lead to the possibility of extender not seating properly on the implant, which may lead to micro-leakage and a possible infection to the patient.

Source facts

Action
Customer notification letters were sent on 12/12/16. Customers were requested to review inventory for the affected product. Complete and return the Acknowledgement and Recall Return Form within 48 hours. If customers are an authorized distributor the firm requests that they identify those customers that may have been shipped the affected product lot and contact those customers to inform them of the issue within 48 hours of receipt of the notification in order to provide the customers with replacement product. If customers have any of the affected product the firm will send a replacement part. Any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425.
Initiated
2016-12-12
Root cause
Packaging process control
Terminated
2017-05-22
Distribution
U.S. distribution to the following; NM, TX, CA, OR, OH, ME. Foreign distribution to the following; Netherlands.
Recall number
Z-1112-2017
Recall status
Terminated
CDRH record ID
152174
FDA classified
2017-01-28
Recall event number
76111

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗