Open original FDA record ↗Recall record
- FDA identifier
- Z-1112-2025
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Feb 5, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
Source facts
- Action
- An Urgent Medical Device Recall letter was sent during the week of December 30, 2024, to the affected customers via first class mail and email. If a passing calibration curve can be generated using the Access PCT reagent lots listed above, no action is needed. If repeated calibration failures with a failure code of Max Iterate are obtained using the Access PCT reagent lots listed above, discontinue using the affected reagent lot number.
- Initiated
- 2025-01-02
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.
- Recall number
- Z-1112-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211985
- FDA classified
- 2025-02-05
- Recall event number
- 96089