Open original FDA record ↗Recall record
- FDA identifier
- Z-1113-2014
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Mar 3, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions used for contrast study may not start.
Source facts
- Action
- Toshiba sent an Urgent Medical Device Correction letter on October 1, 2013, via USPS return receipt mail and accompanied by a response form. The recall notification letter provided the Device Model, serial number, SID, problem, corrective action, request to customers, and contact information (800) 421-1968 or (800) 521-1968. Toshiba designed and test an upgraded ROM for the SC2 PWB to prevent recurrence of the fault. Service technicians will replace the ROM on site at no cost to the system owners. All repairs are expected to be completed by April 2014 and progress reports will be provided on the effectiveness of the plan. For questions regarding this recall call 714-730-5000.
- Initiated
- 2013-10-01
- Root cause
- Software design
- Terminated
- 2015-03-03
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1113-2014
- Recall status
- Terminated
- CDRH record ID
- 122559
- FDA classified
- 2014-03-03
- Recall event number
- 66527