Z-1113-2025: Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Produ

Recall record Z-1113-2025 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1113-2025
Company in FDA record
Sysmex America, Inc.
FDA classification date
Feb 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Source facts

Action
A Product Notification Field Correction letter was sent to customers beginning 1/6/25. Actions: 1. Until a Sysmex representative contacts you, please do not use the instrument running PS-10 software version 1.5 or 1.6 unless it is used only to create cocktails. 2. Review the content of this communication with your facility's physician and/or pathologist and retain this letter for any future reference. 3. Please fill out the form below and email it back to Sysmex using the email address: FlowPS10@Sysmex.com Questions If you have any questions concerning this information, please contact the Technical Assistance Center at 1- 888-879-7639 An additional Product Notification Field Correction letter was sent to customers beginning 4/9/25. Software Correction The probe washing processes have been updated to incorporate essential washing steps during antibody pipetting. The PS-10 antibody pipetting washing sequence has been reinstated, ensuring a washing step is performed when switching between antibody vials. This update prevents any potential carry-over between antibody reagents. This correction addresses the potential risk of antibody carryover, as detailed in Field Correction Notice CF-08823, released in December 2024. Action 1. Please distribute this product notification as appropriate to your laboratory staff. 2. File this product notification as a part of your laboratory s quality system as required. Questions If you have questions concerning this information, please message the Sysmex Technical Assistance Center (TAC). Simply click on the support tab on the home page of the CRC, click the Create New Issue button under Help Me to send your question to TAC. Choose Technical Assistance-Equipment in the dropdown to complete the form and submit it. Sending a message to TAC is recommended for non-urgent requests. For Canadian customers, please note messages to TAC are only processed in English. For urgent requests, contact the Technical Assistance Center at 1-8
Initiated
2025-01-06
Root cause
Software design
Distribution
US Nationwide.
Recall number
Z-1113-2025
Recall status
Open, Classified
CDRH record ID
212016
FDA classified
2025-02-05
Recall event number
96107

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗