- FDA identifier
- Z-1114-2013
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Apr 15, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
Source facts
- Action
- Consignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a day, seven days a week at 1-800-440-3638 for Accu check questions and 1-800-428-4674 for CoaguChek questions about the information contained in this UMDC.
- Initiated
- 2013-03-06
- Root cause
- Device Design
- Terminated
- 2018-01-10
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1114-2013
- Recall status
- Terminated
- CDRH record ID
- 116490
- FDA classified
- 2013-04-15
- Recall event number
- 64460