Open original FDA record ↗Recall record
- FDA identifier
- Z-1114-2025
- Company in FDA record
- US Endodontics, LLC
- FDA classification date
- Feb 6, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to devices being irradiated above the established specification which my result in package integrity issues.
Source facts
- Action
- On 01/02/2025, the firm email an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to customer informing them that sterility of the product may have been compromised due to over irradiation during routine sterilization processing. Customers are instructed to: 1. Confirm if they have any of the impacted Lots. 2. if they have any of the impacted Lots, they are to evaluate the package seal for any potential seal breaches or degradation of the packaging material, which may indicate a lack of sterility in the product. 3. If there are any concerns with the package seal integrity, treat the device as non-sterile product and follow the sterilization instructions in DES-DFU-21001 Rev. F page 6 which provides the following instructions: . Files are single use only and not meant for reprocessing. . Autoclaving should be performed immediately before use. . Place the instruments unwrapped in an autoclave tray. . Use fresh distilled or deionized water. . Insert in a steam gravity cycle autoclave at 134C-137C with a max temp of 140C for a minimum 3 minutes. . Aseptic transport to the point of use should follow autoclaving. . Storage of the sterilized device is not recommended. For Questions contact US Endodontics Customer Care Service at +1 423-722-5626.
- Initiated
- 2025-01-02
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the state of GA and the countries of Australia, Canada, the European Union (specifically France, Germany, Italy, Poland, and Spain), and the UK.
- Recall number
- Z-1114-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211175
- FDA classified
- 2025-02-06
- Recall event number
- 95776