Z-1114-2025: Brasseler ESR Rotary/Reciprocating Endodontic File Device SKU AP120213PK AP120253PK AP120313PK AP125213P

Recall record Z-1114-2025 for US Endodontics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1114-2025
Company in FDA record
US Endodontics, LLC
FDA classification date
Feb 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to devices being irradiated above the established specification which my result in package integrity issues.

Source facts

Action
On 01/02/2025, the firm email an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to customer informing them that sterility of the product may have been compromised due to over irradiation during routine sterilization processing. Customers are instructed to: 1. Confirm if they have any of the impacted Lots. 2. if they have any of the impacted Lots, they are to evaluate the package seal for any potential seal breaches or degradation of the packaging material, which may indicate a lack of sterility in the product. 3. If there are any concerns with the package seal integrity, treat the device as non-sterile product and follow the sterilization instructions in DES-DFU-21001 Rev. F page 6 which provides the following instructions: . Files are single use only and not meant for reprocessing. . Autoclaving should be performed immediately before use. . Place the instruments unwrapped in an autoclave tray. . Use fresh distilled or deionized water. . Insert in a steam gravity cycle autoclave at 134C-137C with a max temp of 140C for a minimum 3 minutes. . Aseptic transport to the point of use should follow autoclaving. . Storage of the sterilized device is not recommended. For Questions contact US Endodontics Customer Care Service at +1 423-722-5626.
Initiated
2025-01-02
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the state of GA and the countries of Australia, Canada, the European Union (specifically France, Germany, Italy, Poland, and Spain), and the UK.
Recall number
Z-1114-2025
Recall status
Open, Classified
CDRH record ID
211175
FDA classified
2025-02-06
Recall event number
95776

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗